Pressure Sensor - GE AKTA start Maintenance Manual

Hide thumbs Also See for AKTA start:
Table of Contents

Advertisement

3.12 Pressure sensor

Parameter description
Parameter
P set
Instructions
Follow the instructions below to set the pressure to 0 when the Pressure sensor is ex-
posed to atmospheric pressure only.
Step
1
ÄKTA start Maintenance Manual 29060308 AD
Description
Displays the current pressure in the flow path (MPa).
NOTICE
Calibration of Pressure sensor must be performed by a GE Service
Engineer, and must be done when the Pressure sensor has been
replaced.
Action
Disconnect the inlet tubing from the Pressure sensor to expose the sensor
to atmospheric pressure only.
Note:
Make sure that the Pump is OFF before disconnecting the tubing.
3 Maintenance operations

3.12 Pressure sensor

75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents