4.7

Auto Step function

Auto Step is used to step automatically between a range of different calibration
temperatures. This is useful when calibrating sensors in places which are hard to
reach, and when calibrating sensors for which the output is displayed in a different
location.
Note...
Temperature range is limited by Min. SET Temp. and Max. SET Temp. settings
editable in the System Settings menu and by the temperature range of the
external reference sensor, if connected.
4.7.1

Running an Auto Step test

F3
1. Press
the Main screen press
2. Press one of the
No of steps: the number of temperature steps per direction (T
Hold time:
Step values: must be set within the sensors permitted range.
129591 03
(Enter) to access the editable field and use the arrows
(Enter) to accept the new set value.
(Back) to return to the previous menu.
(Auto Step) to access the Auto Step setup. If you wish to return to
F1
(Main).
(Arrow) keys to access the editable fields for new values:
set using integers from 2 to 12. You must select minimum 2
steps, maximum 12 steps.
defines the time (in minutes) the temperature is maintained
(after it is stable) for each step.
2019-01-30
(Up) and
T
) can be
1
x
19