4-20 Ma Output Wiring - Ametek 3050-OLV User Manual

Moisture analyzer
Hide thumbs Also See for 3050-OLV:
Table of Contents

Advertisement

4–20 mA Output Wiring
Figure 2-6.
3050-OLV Analyzer Power and
Signal Connections.
PN 305200901, Rev YT
4-20 mA Output, Loop Powered (TB-2)
3
4
4-20 mA Output, Self Powered (TB-2)
3
4
Notes
1. Cable should be shielded with single twisted pair.
The cable should be shielded with single twisted pair.
2. Cable shields should be connected to both the analyzer and the DCS. If this is not
possible, cable shields should be tied to the chassis at each 3050-OLV. If this is not
possible, tie the shield at the PC or DCS to chassis and remaining shield to the chassis
through a 0.1 mF @ 500V capacitor
Cable shields should be connected to both the analyzer and the DCS.
3. The 3050-OLV signal common is connected to earth ground. If the analog output is also
If this is not possible, cable shields should be tied to the chassis at each
grounded, the analog output will no longer be electrically isolated. Contact AMETEK if
3050-OLV. If this is not possible, tie the shield at the PC or DCS to chassis
this situation occurs.
and remaining shield to the chassis through a 0.1 mF @ 500 V capacitor.
Analyzer power must be removed when connecting or disconnecting the 4-20 mA signal.
The 4-20 mA loop circuit must have a load resistance of between 100 and 500 ohms or
malfunction may occur. If a loop check is performed, the resistor must be placed in series
The 3050-OLV Analyzer signal common is connected to earth ground.
with the ohmmeter.
If the analog output is also grounded, the analog output will no lon-
ger be electrically isolated. Contact AMETEK if this situation occurs.
Analyzer power must be removed when connecting or disconnecting
the 4–20 mA signal.
The 4–20 mA loop circuit must have a load resistance of between
100–500 ohms or malfunction may occur. If a loop check is performed,
a 100 ohm resistor must be placed in series with the ammeter.

4-20 mA Output Wiring

5
6
+
5
6
100 to 500 W
+
R
Installation and Start-Up
100 to 500 W
-
R
Load
+ -
External 24V DC
Supply
-
Load
|
2-13