2.4.2 Electromagnetic Emission

The Innova system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Innova system should assure that it is used in an electromagnetic
environment as described below:
Table 3-7:
Emissions
Radio–Frequency
Group1 Class A
Emissions
limits
CISPR11
Group1 Class A
limits
Not applicable
Harmonic emissions
IEC 61000–3–2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000–3–3

2.4.3 Electromagnetic Immunity

2.4.3.1 Electromagnetic Immunity
The Innova system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Innova system should assure that it is used in an electromagnetic
environment as described below:
Table 3-8:
Immunity Test
IEC 60601–1–2 Test
Electrostatic
6 kV contact
discharge (ESD)
8 kV air
IEC 61000–4–2
52
Test Compliance
The Innova system is suitable for use in all establishments other than domestic and
those directly connected to the low voltage power supply network that supplies buildings
used for domestic purposes.
The Innova system uses RF energy only for its internal function. Therefore, the RF
emission is very low and not likely to cause any interference in nearby electronic
equipment.
The Innova system is suitable for use only in establishments not directly connected to a
public low voltage power supply network.
The Innova system is suitable for use only in establishments not directly connected to a
public low voltage power supply network.
Compliance Level
Level
6 kV contact
8 kV air
Direction 5314528-2-1EN, Revision 1
Electromagnetic Environment
Electromagnetic Environment
Floors are wood, concrete, or ceramic tile, or floors are covered with
synthetic material and the relative humidity is at least 30 percent.
2 Environmental Requirements/Limitations