Configure The Backlight Settings; Configure The Digital Output Alarms - Hach PAT700 User Manual

Toc analyzer
Hide thumbs Also See for PAT700:
Table of Contents

Advertisement

4. Select the Network tab.
5. Select an option.
Option
Static TCP/IP
Address
Static Subnet Mask
Static Gateway
DNS Server
WINS Server
DHCP

Configure the backlight settings

1. Push << to open the sliding menu.
2. Push the SETUP icon.
3. Push SYSTEM.
4. Select the Backlight tab.
5. Configure the backlight options.
Option
Automatically turn off backlight Have the backlight switch off automatically after a period of inactivity.
Turn off after
6. To adjust the brightness of the backlight, push the brightness slider.

Configure the digital output alarms

Assign output functionally for each digital output as necessary. Refer to
outputs, refer to
Configure digital control
Note: The TOC alarm can be assigned to multiple digital outputs (1–4) at the same time to set a multi-level alarm
on the TOC result.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push ALARMS.
4. Assign the output functionality for each digital output. Refer to
Note: Conductivity alarm, Error state and Valve state can only be assigned to one digital output at a time.
5. Select the checkbox to enable the alarm, if present.
6. For TOC alarm, select the upper TOC limit for the alarm. Options: 1 to 2500 (default = 500). This
value applies only to online TOC analysis, not to bottle or grab sample analysis.
Description
Set the TCP/IP address for the analyzer. Options: 0 to 255 for each octet (default =
192.168.0.105)
Set the subnet mask for the analyzer. Options: 0 to 255 for each octet (default =
255.255.255.0)
Set the default gateway for the analyzer. Options: 0 to 255 for each octet (default =
blank)
Set the DNS server address for the analyzer. Options: 0 to 255 for each octet
(default = blank)
Set the WINS server address for the analyzer. Options: 0 to 255 for each octet
(default = blank)
Enable DHCP. When selected, the TCP/IP address, subnet mask and default
gateway entries are disabled (default = enabled).
Description
Select the number of minutes of inactivity before the backlight switches
off. Options: 2, 5, 10, 15 or 30 minutes
on page 28.
Table
13. To invert the digital
Table
13.
English 31