9705 - Diagnostics and Troubleshooting

6
Diagnostics and Troubleshooting

Error Messages

6.1
When error messages occur, a pop-up window appears with the error message highlighted.
Any sound effects related to errors will be heard.
• Press Silence Alarm to stop the sound but keep the error message on the screen.
• Press OK to close the error message pop-up window and view the Diagnostics tab.
Error messages are described in
Error Messages
Table 6-1 :
Error Message
Syringe door open
Communication error
Flow system error

Troubleshooting Steps

6.2
Table 6-2
9705 Symptoms and Solutions
Table 6-2 :
Symptom
Unit will not power up.
Screen is dark.

Diagnostics Screen

6.3
HIAC
Cause
Occurs when the syringe door is ajar
Occurs when communications are lost
Occurs when there is a problem with the flow
of sample through the 9705
shows a list of symptoms, possible causes and solutions for various 9705 problems.
Possible Cause
Unit may be disconnected from AC
power.
Check fuses in AC power cord.
Backlight timeout is in effect.

Fig 6-1 : Diagnostics Icon

Table
6-1.
Solutions
Close syringe door
• Double-check all communications connec-
tions.
• Reboot the 9705.
• Double-check all communications connec-
tions.
• Reboot the 9705.
Solution
Check AC power connection.
Replace faulty fuses.
Tap touchscreen.
If tone sounds and screen still does not
light, contact technical support.
65 of 82
Operator Manual