Operating in Electron Ionization (EI) Mode

LCP status messages

The following messages may be displayed on the LCP to inform you of the
status of the MSD system. If the LCP is currently in Menu mode, cycle through
the menus to return to Status mode. No messages will be displayed if an online
instrument session is not currently running in MassHunter Data Acquisition.
ChemStation Loading <timestamp>
The Agilent MassHunter Data Acquisition software is starting up.
Executing <type>tune
A tuning procedure is in progress (type = QuickTune or Autotune).
Instrument Available <timestamp>
The Agilent MassHunter Data Acquisition software is not running.
Loading Method <method name>
Method parameters are being sent to the MSD.
Loading MSD Firmware
The MSD's firmware is being initialized.
The following messages alternately appear on the LCP if the MSD does NOT
complete its bootup sequence properly:
These messages indicate that the MSD has not received its unique IP address
from the Windows Service. If the messages persist after you have logged onto
your account in the MassHunter Data Acquisition program, consult the
Troubleshooting section of the Software Installation manual.
Loading OS
The operating system of the instrument controller is being initialized.
48
Server not Found
Check LAN Connection
Seeking Server
Bootp Query xxx
5977 Series MSD Operation Manual