To Install the CI/Extractor Interface Tip Seal

Materials needed
• Interface tip seal (G1999-60412)
Electrostatic discharges to analyzer components are conducted to the side board
CA UT IO N
where they can damage sensitive components. Wear a grounded antistatic wrist strap
and take other antistatic precautions before you open the analyzer chamber.
Procedure
1
2
CI/Extractor tip seal
5977 Series MSD Operation Manual
The interface tip seal must be in place for the CI ion source and the
Extraction source.
Verify that the EI extractor ion source or CI ion source is installed. This tip
seal must not be installed when a standard EI SST source or EI Inert
source is installed.
Remove the CI/Extractor tip seal from the ion source storage box and place
it over the end of the interface.