6
CI Maintenance
Table 28
Wire color
Blue
Orange
White
Red
Black
4
Pull on the connectors, not on the wires.
CA UT IO N
5
6
The analyzer operates at high temperatures. Do not touch any part until you are sure
WARNING
it is cool.
158
Standard CI ion source wires
Connects to
Entrance lens
Ion focus
Filament 1 (top
filament)
Repeller
Filament 2 (bottom
filament)
Trace the wires for the ion source heater and temperature sensor to the
feedthrough board. Use the pliers to pull on the metal connectors to
remove these four wires from the feedthrough board connections.
Remove the thumbscrews that hold the ion source in place.
Pull the ion source out of the source radiator.
Number of leads
1
1
2
1
2
5977 Series MSD Operation Manual