Agilent Technologies G7035A Operation Manual page 51

5977 series msd
Table of Contents

Advertisement

Table 7
Action
Prepare to vent
Pumpdown
Hi Vac Soft Start
Table 8
Action
High Vacuum Pressure
Turbo Pump Speed
MSD Fault Status
Ion Source Temp,
Quadrupole Temp,
CI Reagent
MS parameters cannot be set from the LCP while an online MassHunter Data Acquisition
NO T E
session is connected to the MSD.
Table 9
Action
MSD IP via keyboard
Gateway IP Address
Subnet Mask
GC Comm IP Address
5977 Series MSD Operation Manual
Maintenance menu
Description
Reminds you to shut down the GC then prepares the instrument for
venting when [Yes/Select] is pressed.
Initiates a pumpdown sequence.
Allows you to connect or disconnect the hi vac soft start feature.
This feature slowly ramps up the turbo pump speed, and should be
used after the pump has sat idle for a few weeks to minimize wear
on the pump.
MS Parameters menu
Description
Only with Micro-Ion vacuum gauge installed.
Displays the turbo pump speed.
Reports a summary fault status code (number) in 'dec' (decimal) and
'hex' (hexadecimal) format covering all possible fault combinations.
o
C
Displays and sets the ion source temperature.
o
C
Displays and sets the quadrupole temperature.
Displays CI reagent gas and flow rate (if installed).
Network menu
Description
Displays the IP address for the MSD and is programmed via
MassHunter Data Acquisition.
Displays the gateway IP address for the MSD.
Displays the subnet mask for the MSD.
Displays the IP address
Operating in Electron Ionization (EI) Mode
3
51