General Maintenance

Maintaining the Analyzer

Scheduling
None of the analyzer components require periodic maintenance. Some tasks,
however, must be performed when MSD behavior indicates they are necessary.
These tasks include:
• Cleaning the ion source
• Replacing filaments
• Replacing the electron multiplier horn
The Agilent 5977 Series MSD Troubleshooting and Maintenance Manual
provides information about symptoms that indicate the need for analyzer
maintenance. The troubleshooting material in the online help in the
MassHunter software provides more extensive information.
Precautions
Cleanliness
Keep components clean during analyzer maintenance. Analyzer maintenance
involves opening the analyzer chamber and removing parts from the analyzer.
During analyzer maintenance procedures, take care to avoid contaminating
the analyzer or interior of the analyzer chamber. Wear clean gloves during all
analyzer maintenance procedures. After cleaning, parts must be thoroughly
baked out before they are reinstalled. After cleaning, analyzer parts should be
placed only on clean, lint-free cloths.
If not done correctly, analyzer maintenance can introduce contaminants into the MSD.
CA UT IO N
The analyzer operates at high temperatures. Do not touch any part until you are sure
WARNING
it is cool.