To Remove the EI Ion Source

Materials needed
• Gloves, clean, lint-free
• Pliers, long-nose (8710-1094)
Procedure
1
2
3
Table 22
Wire color
Blue
Orange
White
Red
Black
Pull on the connectors, not on the wires.
CA UT IO N
128
Large (8650-0030)
Small (8650-0029)
Vent the MSD. See
"To Vent the MSD"
Open the analyzer chamber. See
page 78.
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
If you are using a standard or inert EI ion source, disconnect the seven
wires from the ion source. Do not bend the wires any more than necessary
(See
Figure 20
on page 130 and
Standard EI ion source wires
Connects to
Entrance lens
Ion focus
Filament 1 (top filament)
Repeller
Filament 2 (bottom filament)
on page 76.
"To Open the Analyzer Chamber"
Table
22).
Number of leads
1
1
2
1
2
5977 Series MSD Operation Manual