Agilent Technologies G7035A Operation Manual page 79

5977 series msd
Table of Contents

Advertisement

3
In the next step, if you feel resistance, stop. Do not try to force the side plate open.
CA UT IO N
Verify that MSD is vented. Verify that both the front and rear side plate screws are
completely loose.
The analyzer, GC/MSD interface, and other components in the analyzer chamber
WARNING
operate at very high temperatures. Do not touch any part until you are sure it is cool.
Always wear clean gloves to prevent contamination when working in the analyzer
CA UT IO N
chamber.
4
Detector
Side plate
Ion source
Feedthrough
board
5977 Series MSD Operation Manual
Loosen the side plate thumbscrews if they are fastened.
The rear side plate thumbscrew should be unfastened during normal use.
It is only fastened during shipping. The front side plate thumbscrew should
only be fastened for CI operation or if hydrogen or other flammable or
toxic substances are used for carrier gas.
Gently swing the side plate out.
Operating in Electron Ionization (EI) Mode
3
79