5
General Maintenance
To Install the EI Ion Source
Materials needed
• Gloves, clean, lint-free
• Pliers, long-nose (8710-1094)
Procedure
1
Ion source
Source feedthrough board
Thumbscrews
Extractor wire
Source heater and temperature
sensor wires
Source radiator
Figure 27
Installing the EI ion source
2
3
4
150
•
Large (8650-0030)
•
Small (8650-0029)
Slide the ion source into the source radiator.
Install and hand tighten the source thumbscrews. Do not overtighten the
thumbscrews.
Connect the ion source wires as shown in
Close the analyzer chamber. See
page 80.
Figure 14
on page 82.
"To Close the Analyzer Chamber"
5977 Series MSD Operation Manual