To Perform High-Mass Testing (5977 Series Msds) - Agilent Technologies G7035A Operation Manual

5977 series msd
Table of Contents

Advertisement

3
Operating in Electron Ionization (EI) Mode

To Perform High-Mass Testing (5977 Series MSDs)

Materials needed
• PFHT calibration sample (5188-5357)
Procedure
1
2
3
4
5
6
7
72
Load tune file ATUNE.U then auto tune the MSD. See
EI Mode"
on page 69.
Resolve the PFHT.M method under x\5977\PFHT.M where x is the
instrument number being used.
Update and save the method.
Load the PFHT calibration sample into a vial and place in position 2.
In Instrument Control view select High Mass Check from the Checkout menu.
Follow the instructions on screen.
The Run is completed and results are printed within 5 minutes. See
Figure 12
on page 73.
"To Tune the MSD in
5977 Series MSD Operation Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G7036aG7038aG7039aG7040aG7037a

Table of Contents