To View Msd Temperature And Vacuum In Manual Tune - Agilent Technologies G7035A Operation Manual

5977 series msd
Table of Contents

Advertisement

3
Operating in Electron Ionization (EI) Mode

To View MSD Temperature and Vacuum in Manual Tune

You can also use the Local Control Panel to perform this task. See
the MSD from the LCP"
Procedure
1
2
3
Unless you have just begun the pumpdown process, the foreline pressure
should be less than 300 mTorr, or the turbo pump should be running at least
80% speed. MSD heaters remain off as long as the diffusion pump is cold or the
turbo pump is operating at less than 80%. Normally, the foreline pressure will
be below 100 mTorr, or the turbo pump speed will be at 100%.
The MSD heaters turn on at the end of the pumpdown cycle and turn off at the
beginning of the vent cycle. The reported setpoints will not change during
venting or pumpdown, even though both the MSD zones are turned off.
58
In Instrument Control view, select Edit Tune Parameters from the Instrument
menu to display the Manual Tune dialog.
Click the Values tab to view the MSD Temperatures and Vacuum.
To change a temperature Setpoint or Limit enter the new parameters and
click Apply.
on page 46.