Agilent Technologies G7035A Operation Manual page 169

5977 series msd
Table of Contents

Advertisement

5
6
7
8
9
10 Perform a PCI autotune with methane. See
11 Select Save Tune Parameters from the File menu.
5977 Series MSD Operation Manual
Secure the new filament with the gold plated screw and washer.
After installing the filament, verify that it is not grounded to source body.
Install the ion source. See
Close the analyzer chamber. See
page 80.
Pump down the MSD. See
page 98.
(Methane Only)"
on page 111.
"To Install the CI Ion Source"
"To Close the Analyzer Chamber"
"To Pump Down the MSD in CI Mode"
"To Perform a PCI Autotune
6
CI Maintenance
on page 167.
on
on
169