Installing GC Columns
6
Figure 9
7
Figure 10
44
Check the end of the tube with a magnifier. The end of the tube need not be
perfectly square, but should not have cracks which extend under the
ferrule.
Figure 9
shows a completed tube end.
Tube end with internal nut and swaged SilTite ferrule
Connect column to the Quick Swap or PUU with internal nuts and
preswaged SilTite ferrules
tighten with a wrench only 15 to 20 degrees.
Completed Ultimate union with columns
(Figure
10). Finger-tighten the nuts. Further
5977 Series MSD Operation Manual