Agilent Technologies G7035A Operation Manual page 67

5977 series msd
Table of Contents

Advertisement

9
10 Verify that the parameters listed (temperature, inlet and outlet pressures,
11 Enter the recorded retention time in the Holdup Time field. Move the cursor
12 Click OK to save the changes and exit the dialog.
13 Click OK on the Calibrate Columns dialog to save the calibration.
With capillary columns, such as those used with the MSD, linear velocity is
often measured rather than volumetric flow rate.
5977 Series MSD Operation Manual
Click the Calc Length button in the If unretained peak holdup time is known
section to display the Calculate Column Length dialog.
and gas type) are those used in the method to determine the holdup time.
Change any parameters that are different than those used in your method.
to another parameter's field and the calibrated column length appears.
Operating in Electron Ionization (EI) Mode
3
67