CI Maintenance
3
Forcing the analyzer closed if these parts are misaligned will damage the seal or the
CA UT IO N
interface or the ion source, or keep the sideplate from sealing.
4
156
Gently check the alignment of the analyzer and the interface.
When the analyzer is aligned correctly, the analyzer can be closed all the
way with no resistance except the spring tension from the interface tip
seal.
You can align the analyzer and interface by wiggling the side plate on its
hinge. If the analyzer still will not close, contact your Agilent Technologies
service representative.
5977 Series MSD Operation Manual