2
Installing GC Columns
To Install a Capillary Column in the GC/MS Interface
This procedure is for the installation of a capillary column directly to the
transfer line with a column nut. If you are using the Agilent capillary flow
technology Quick Swap accessory, or a Purged Ultimate Union (PUU) see
Prepare the Column Ends for a CFT Fitting"
Agilent 7890 Series GC
Materials needed
• Column cutter, ceramic (5181-8836) or diamond (5183-4620)
• Ferrules
• Flashlight
• Magnifying loupe
• Gloves, clean
• Interface column nut (05988-20066)
• Safety glasses
• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
• Column measuring tool
Always wear clean gloves while handling any parts that go inside the GC or the
CA UT IO N
analyzer chambers.
38
•
0.3 mm id, for 0.10 mm id columns (5062-3507)
•
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
•
0.5 mm id, for 0.32 mm id columns (5062-3506)
•
0.8 mm id, for 0.53 mm id columns (5062-3512)
•
Large (8650-0030)
•
Small (8650-0029)
on page 41.