If you are using an extractor EI ion source, disconnect the eight wires from
the ion source. Do not bend the wires any more than necessary (See
Figure 20
on page 130 and
Extractor EI ion source wires
Connects to
Entrance lens
Ion focus
Filament 1 (top filament)
Repeller
Filament 2 (bottom filament)
Extractor lens
Trace the wires for the ion source heater and temperature sensor to the
feedthrough board. Disconnect them there.
Remove the thumbscrews that hold the ion source in place.
Pull the ion source out of the source radiator.
Table
23).
Number of leads
1
1
2
1
2
1