Agilent Technologies G7035A Operation Manual page 129

5977 series msd
Table of Contents

Advertisement

4
Table 23
Wire color
Blue
Orange
White
Red
Black
Brown
5
6
7
The analyzer operates at high temperatures. Do not touch any part until you are sure
WARNING
it is cool.
5977 Series MSD Operation Manual
If you are using an extractor EI ion source, disconnect the eight wires from
the ion source. Do not bend the wires any more than necessary (See
Figure 20
on page 130 and
Extractor EI ion source wires
Connects to
Entrance lens
Ion focus
Filament 1 (top filament)
Repeller
Filament 2 (bottom filament)
Extractor lens
Trace the wires for the ion source heater and temperature sensor to the
feedthrough board. Disconnect them there.
Remove the thumbscrews that hold the ion source in place.
Pull the ion source out of the source radiator.
Table
23).
Number of leads
1
1
2
1
2
1
General Maintenance
5
129

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G7036aG7038aG7039aG7040aG7037a

Table of Contents