N/A Not Available - Agilent Technologies G7035A Operation Manual

5977 series msd
Table of Contents

Advertisement

4
Operating in Chemical Ionization (CI) Mode
Table 15
Reagent gas settings
Reagent gas
Ion polarity
Positive
150 A
Emission
Electron
150 eV
energy
Filament
1
Repeller
3 V
Ion focus
130 V
Entrance
20 V
lens offset
EM volts
1200
Shutoff valve Open
Gas select
A
Suggested
20%
flow
Source temp
250 °C
Quad temp
150 °C
Interface
280 °C
temp
Autotune
Yes
94
Methane
Negative
Positive
50 A
150 A
150 eV
150 eV
1
1
3 V
3 V
130 V
130 V
20 V
20 V
1400
1200
Open
Open
A
B
40%
20%
150 °C
250 °C
150 °C
150 °C
280 °C
280 °C
Yes
No

N/A Not available

Isobutane
Negative
Positive
50 A
150 A
150 eV
150 eV
1
1
3 V
3 V
130 V
130 V
20 V
20 V
1400
1200
Open
Open
B
B
40%
20%
150 °C
250 °C
150 °C
150 °C
280 °C
280 °C
Yes
No
Ammonia
EI
Negative
N/A
50 A
35 A
150 eV
70 eV
1
1 or 2
3 V
30 V
130 V
90 V
20 V
25 V
1400
1300
Open
Closed
B
None
40%
N/A
150 °C
230 °C
150 °C
150 °C
280 °C
280 °C
Yes
Yes
5977 Series MSD Operation Manual