Important Safety Warnings - Agilent Technologies G7035A Operation Manual

5977 series msd
Table of Contents

Advertisement

Important Safety Warnings

There are several important safety notices to always keep in mind when using
the MSD.
Many internal parts of the MSD carry dangerous voltages
If the MSD is connected to a power source, even if the power switch is off,
potentially dangerous voltages exist on:
• The wiring between the MSD power cord and the AC power supply, the AC
With the power switch on, potentially dangerous voltages also exist on:
• All electronics boards in the instrument
• The internal wires and cables connected to these boards
• The wires for any heater (oven, detector, inlet, or valve box)
All these parts are shielded by covers. With the covers in place, it should be difficult
WARNING
to accidentally make contact with dangerous voltages. Unless specifically
instructed to, never remove a cover unless the detector, inlet, or oven are turned off.
If the power cord insulation is frayed or worn, the cord must be replaced. Contact
WARNING
your Agilent service representative.
If one of the primary fuses has failed, the MSD will already be off, but for
safety you should switch off the MSD and unplug the power cord. It is not
necessary to allow air into the analyzer chamber.
Never replace the primary fuses while the MSD is connected to a power source.
WARNING
Electrostatic discharge is a threat to MSD electronics
The printed circuit boards in the MSD can be damaged by electrostatic
discharge. Do not touch any of the boards unless it is absolutely necessary. If
you must handle them, wear a grounded wrist strap and take other antistatic
precautions.
5977 Series MSD Operation Manual
power supply itself, and the wiring from the AC power supply to the power
switch.
1
Introduction
17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G7036aG7038aG7039aG7040aG7037a

Table of Contents