Prepare the Logic Analysis System for Testing

1 Record the 16910A or 16911A logic analyzer's model and serial number in
the Performance Test Record (see page 60). Record your work order
number (if applicable) and today's date.
2 Record the test equipment information in the "Test Equipment Used"
section of the Performance Test Record.
3 Turn on the logic analysis system.
a Connect the keyboard and monitor to the rear panel of the logic
analysis mainframe (16900A only).
b Connect the mouse to the rear panel of the mainframe.
c Plug in the power cord to the power connector on the rear panel of the
mainframe.
d Turn on the main power switch on the mainframe front panel.
While the logic analysis system is booting, observe the boot dialog for the following:
• ensure all of the installed memory is recognized
• any error messages
• interrupt of the boot process with or without error message
4 During initialization, check for any failures.
If an error or an interrupt occurs, refer to the Agilent Technologies 16900A-series
Logic Analysis System Service Guide for troubleshooting information.
Chapter 3: Testing Logic Analyzer Performance
Prepare the Logic Analysis System for Testing