To Adjust Adc - Agilent Technologies 89410A Service Manual

Vector signal analyzer
Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

Adjusting the Analyzer

To adjust ADC

To adjust ADC
This procedure adjusts the A21 A/D Converter assembly's ADC. The
Agilent 89410A optimizes the ADC circuits. This procedure only needs to
be done if an A21 A/D Converter assembly is replaced.
Equipment Required:
None
1
2
3
This completes the A21 A/D Converter assembly adjustment.
2-24
Set the power switch to on ( l ).
Press the following keys:
]
[ System Utility
]
[ more
[
diagnostics
]
[
service functions
]
1125
[
]
enter
[
]
adjustments
[
ADC settle dither
]
Wait until the Agilent 89410A completes the adjustment.
It takes about 6 minutes to perform the ADC software adjustment.
The message box will disappear when the adjustment is done.