Adjusting the Analyzer
To adjust 10 MHz low pass filter
[
analog source
]
[
]
setup 1
3
Set up the network analyzer.
a
Set the network analyzer as follows:
Receiver
Attenuation
Resolution BW
Display format
Source
Sweep Mode
Sweep Type
Sweep Time
Amplitude
Start Frequency
Stop Frequency
b
Connect the network analyzer's reference channel to the
Agilent 89410A's source using an N(m)-to-BNC(f) adapter and
BNC cable.
c
Connect the network analyzer's source to A35 TP14 (AMPIN) and
TP13 (Ground) using an N(m)-to-BNC(f) adapter, BNC cable, and
BNC(f)-to-test clips adapter.
2-26
reference channel
20 dB
100 Hz
log magnitude
continuous
linear frequency sweep
3 s
-40 dBm
10 MHz
30 MHz