Agilent Technologies 89410A Service Manual page 143

Vector signal analyzer
Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

4
Normalize the measurement.
a
Wait for a complete sweep on the network analyzer.
b
Store the trace data in a storage register.
c
Set the network analyzer to divide the reference channel by the
storage register.
Adjusting the Analyzer
To adjust 10 MHz low pass filter
2-27