To keep the sine wave displayed, adjust the oscilloscope's trigger
level while varying the trigger offset value.
d
If the trigger offset function is incorrect, the A10 Analog Input
assembly is probably faulty.
e
Reconnect A10 J3 to A36 J5.
f
Disconnect the synthesizer from CHANNEL 1.
4
Check the analog source to input path.
a
Press the following keys:
]
[Preset
[
]
Source
[
source on]
]
[level
0
[
dBm
]
[
System Utility
]
[
auto cal off]
]
[more cal setup
[
auto zero cal off]
]
[Return
[
more
]
[
diagnostics
]
[
service functions
[
special test modes
[
]
input
[
source
]
b
Connect the oscilloscope to A10 J2 using the BNC(m)-to-SMB(f)
cable.
c
The oscilloscope should display a 130 ±13 mVp-p sine wave.
d
If the signal is correct, go to Step 5.
e
Place the analyzer on the side closest to the display.
f
Remove the bottom cover.
g
Set the oscilloscope as follows:
CH1 V/div
Input Impedance
CH1 Coupling
]
]
100 mV/div
1 M
dc
Troubleshooting the Analyzer
To troubleshoot input and ADC failures
1-73