Agilent Technologies 89410A Service Manual page 63

Vector signal analyzer
Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

5
Check the HPIB controller on the A42 Memory assembly.
a
Press the following keys:
]
[System Utility
[
]
more
[
diagnostics
]
[
functional tests
]
[
I/O
]
[
]
HPIB controller
b
If the HPIB controller test fails, the A42 Memory assembly is
probably faulty.
6
Check the oven fuse on the A95 Main Power Supply assembly.
a
Set the power switch to off ( O ).
b
Turn the analyzer upside down.
c
Remove the bottom cover.
d
Wait five minutes to allow time for the power supply capacitors to
discharge.
e
Remove the black power supply shield.
f
Measure the resistance across F357.
g
If the resistance is 0
h
If the resistance was greater than 0
, the A85 Oven assembly is probably faulty.
, replace the fuse.
Troubleshooting the Analyzer
To perform self tests
1-45

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents