Advertisement

5
Check the random noise output.
a
b
Oscilloscope Setup
Connect CH1 to SOURCE
CH1 V/div
100 mV/div
50 Ω
Input Impedance
CH1 Coupling
dc
Probe Atten
1
Display Mode
Real Time
Resolution
7 bits
Time/div
200 ns/div
Trigger
Trg'd Sweep
Trig Src
Chan1
c
Press the following keys:
[cal signal off]
]
[Source
[
source on]
]
[level
1
[
]
V
[
]
source type
[
]
random noise
Using an oscilloscope and 1:1 probe, check the following signal.
Parameters
Waveform
Variations
No Pattern
Random Noise
If the random noise source is incorrect, the A30 Digital Source
assembly is probably faulty.
This is only a quick check of the random noise signal. Do the
Random Noise performance test to thoroughly check the random
noise signal.
Troubleshooting the Analyzer
To troubleshoot source and calibrator out failures
1-67