Troubleshooting the Analyzer
To troubleshoot intermittent failures
5
Check the calibration correction curves.
a
Press the following keys:
]
[
Preset
[
]
System Utility
[
]
auto cal off
]
[
more cal setup
[
auto zero cal off]
[
Measurement Data
[
]
freq response
]
[
Ref Lvl/Scale
[
]
Y ref level
0
[
]
dB
[
]
Y per div
0.5
[
]
dB
[
rng tracking off]
]
[
ref position
50
[
]
%
[
]
ref line on
Since there is no input, the display will show lines across the
screen.
b
Press the following keys for each of the primary calibration
ranges:
The primary calibration ranges are -14, −12, −10, −8, −6, and 0
dBm.
]
[
Range
[
]
ch1 range
primary calibration range
]
[
dBm
]
[
System Utility
[
]
more cal setup
[
save ch1 cal trace
[
]
into D1
[
Measurement Data
[
]
data reg
[
]
D1
The following is a typical calibration correction curve display:
1-56
]
]
]