Agilent Technologies 89410A Service Manual page 83

Vector signal analyzer
Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

3
Check the periodic chirp output.
a
b
Oscilloscope Setup
Connect CH1 to SOURCE
CH1 V/div
300 mV/div
50 Ω
Input Impedance
CH1 Coupling
dc
Probe Atten
1
Display Mode
Repetitive
Averaging
off
Persistence
500 ms
Time/div
50 ns/div
Trigger
Trg'd Sweep
Trig Src
Chan1
c
Press the following keys:
]
[DC offset
0
[
V
]
[
source type
]
[
periodic chirp
]
Using an oscilloscope and 1:1 probe, check the following signal.
Parameters
Waveform
Variations
Periodic Chirp
If the signal is incorrect, go to Step 7.
Troubleshooting the Analyzer
To troubleshoot source and calibrator out failures
1-65