Troubleshooting the Analyzer
To troubleshoot source and calibrator out failures
4
Check the calibrator output.
a
b
Oscilloscope Setup
Connect CH1 to SOURCE
CH1 V/div
20 mV/div
50 Ω
Input Impedance
CH1 Coupling
dc
Probe Atten
1
Display Mode
Repetitive
Averaging
8
Time/div
500 ns/div
Trigger
Trg'd Sweep
Trig Src
Chan1
Trigger Level
0 V
c
1-66
Press the following keys:
]
[Preset
[
]
System Utility
[
auto cal off]
[more cal setup
]
[
auto zero cal off]
]
[Return
[
]
more
[
]
diagnostics
[
service functions
]
1125
[
]
enter
[
]
special test modes
[
]
cal
[
cal signal
on
]
Using an oscilloscope and 1:1 probe, check the following signal.
Parameters
Amplitude
Time
Duty Cycle
Pulse Shape
Calibrator
If the signal is incorrect, go to Step 7.
Waveform