Adjusting the Analyzer
To adjust reference oscillator
To adjust reference oscillator
This procedure adjusts the A60 Frequency Reference assembly's
reference oscillator.
Equipment Required:
Frequency counter
Extender board, Agilent part number 89410-B1002
SMB(f)-to-BNC(m) cable
Flat-edge adjustment tool, Agilent part number 8710-1928
Plastic screw driver, Agilent part number 8710-2056
1
2-32
Place the A60 Frequency Reference assembly on an extender board.
a
Set the power switch to off ( O ).
b
Remove screws holding the A60 Frequency Reference assembly in
the card nest.
c
Place the assembly on an extender board.