Troubleshooting the Analyzer
To troubleshoot the power supply
4
Check the A95 Main Power Supply assembly's bias supply voltages.
a
Check the bias supply voltages in the following table.
The ripple should be no more than 50 mVp-p. Do not include high
frequency line noise when measuring ripple.
Test
Location
U341, pin 3
U340, pin 2
U350, pin 3
b
If any of the voltages are incorrect, the A95 Main Power Supply is
probably faulty.
The A95 Main Power Supply assembly has no adjustments.
Therefore if a voltage is incorrect, replace the assembly.
c
If the voltages are correct and the fan is turning, go to Step 7.
5
Check the supply voltage fuse for the A102 Fan assembly.
a
Set the power switch to off ( O ).
b
Wait five minutes to allow time for the power supply capacitors to
discharge.
c
Measure the resistance across A95 F353.
d
If the resistance is greater than 0
1-20
Nominal
Minimum
Voltage
Voltage
+5 V
+4.5 V
+12 V
+11.4 V
−12 V
−12.6 V
, replace the fuse.
Maximum
Ripple
Voltage
Tolerance
+5.5 V
50 mVp-p
+12.6 V
50 mVp-p
−11.4 V
50 mVp-p