Agilent Technologies 89410A Service Manual page 183

Vector signal analyzer
Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

Replacing Assemblies
A100 Display, Backlights, and Filter
Backlight Removal/Replacement
There are two backlight assemblies installed in the flat panel display.
NOTE: The backlight lamps should be replaced as a pair, even if only one
lamp is bad. The other one might fail shortly.
Rthe
Refer to the "Display Removal" procedure on
page
3-27. Unplug the
backlight cables from the backlight connectors
It is not necessary to remove the display mount from the front frame to
change the backlights.
The following Figure shows how the backlights slide out of the display.
Carefully pull on the tab on the backlight assembly (2) to slide the
backlight partially out of the display. Once the end of the backlight casing
is exposed, you can pull it straight out from the display.
NOTE: Do not touch the bulb encased in the plastic backlight assembly.
3-29