Troubleshooting the Analyzer
To troubleshoot the power supply
9
Isolate assembly loading power supply.
a
Set the power switch to off ( O ) and disconnect the power cord
from the rear panel.
b
Wait five minutes to allow time for the power supply capacitors to
discharge.
c
Disconnect A95 P360 pins 2 and 3.
d
Reconnect all the cables to the A95 Main Power Supply assembly.
e
Reinstall the A95 Main Power Supply assembly in the analyzer.
f
Replace the black power supply shield and bottom cover.
g
Repeat the following steps for each assembly suspected of loading
the power supply.
Use the Power Supply Voltage Distribution table on page 6- and
the A90/A91 Motherboard Voltages table on page 6- to help isolate
the failing assembly.
h
Set the power switch to off ( O ).
i
Remove the suspected assembly or the power cable to the
assembly.
j
Set the power switch to on ( l ).
k
Check the voltages in the table on page 1-21.
1-24