Agilent Technologies 89410A Service Manual page 122

Vector signal analyzer
Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

Adjusting the Analyzer
Remote Operation
Adjustments can be setup using the remote operation capability of the
Agilent 89410A analyzer. The following table lists the adjustments and
corresponding HPIB codes. See the Agilent 89400 Series HPIB
Command Reference for general information on remote operation.
Adjustment
Channel 1 Input flatness setup 1
Channel 1 Input flatness setup 2
Channel 1 Input flatness setup 3
Channel 2 Input flatness setup 1
Channel 2 Input flatness setup 2
Channel 2 Input flatness setup 3
Channel 1 Input capacitance setup 1
Channel 1 Input capacitance setup 2
Channel 2 Input capacitance setup 1
Channel 2 Input capacitance setup 2
Channel 1 Input offset
Channel 2 Input offset
Channel 1 Anti-alias filter setup 1
Channel 1 Anti-alias filter setup 2
Channel 1 Anti-alias filter setup 3
Channel 2 Anti-alias filter setup 1
Channel 2 Anti-alias filter setup 2
Channel 2 Anti-alias filter setup 3
ADC
10 MHz low pass filter setup 1
10 MHz low pass filter setup 2
Autorange detect level
Display brightness
2-6
HPIB Code
DIAG:ADJ:INP1:FLAT 1
DIAG:ADJ:INP1:FLAT 2
DIAG:ADJ:INP1:FLAT 3
DIAG:ADJ:INP2:FLAT 1
DIAG:ADJ:INP2:FLAT 2
DIAG:ADJ:INP2:FLAT 3
DIAG:ADJ:INP1:CAP 1
DIAG:ADJ:INP1:CAP 2
DIAG:ADJ:INP2:CAP 1
DIAG:ADJ:INP2:CAP 2
DIAG:ADJ:INP1:FET 1
DIAG:ADJ:INP2:FET 1
DIAG:ADJ:INP1:AAF 1
DIAG:ADJ:INP1:AAF 2
DIAG:ADJ:INP1:AAF 3
DIAG:ADJ:INP2:AAF 1
DIAG:ADJ:INP2:AAF 2
DIAG:ADJ:INP2:AAF 3
DIAG:ADJ:ADC:SDIT
DIAG:ADJ:SOUR 1
DIAG:ADJ:SOUR 2
DIAG:MODE:TRIG:LEV 128
DISP:BRIG 100 PCT