Troubleshooting the Analyzer
To troubleshoot source and calibrator out failures
6
Step 6. Check the source attenuation.
a
Connect a digital multimeter to the SOURCE output using a 50
feedthrough termination and BNC cable.
b
Press the following keys:
[CW (fixed sine)
]
[
Return
]
[
]
sine freq
1
[
]
kHz
[
level
]
5
[
]
V
[
]
System Utility
[
]
more
[
]
diagnostics
[
service functions
[
special test modes
[
]
source
[
]
Marker/Entry
c
Slowly rotate the RPG knob while monitoring the digital
multimeter. The sine wave's amplitude should vary as follows.
Source
Attenuation
0
1
2
3
4
5
6
7
d
If all the values are incorrect, go to Step 7.
e
If only a few of the values are incorrect, the A35 Analog Source is
probably faulty.
f
If all the values are correct, the analog source and calibrator are
operating correctly. For a single channel analyzer, go to
page
1-71, ''To troubleshoot input and ADC failures.'' For a two
1-68
]
]
Source Amplitude (nominal)
dBm
Vp-p
24
10.0
14
3.17
4
1.00
−6
317 m
−16
100 m
−26
31.7 m
−36
10.0 m
−46
3.17 m
Vrms
3.54
1.12
354 m
112 m
35.4 m
11.2 m
3.54 m
1.12 m