Agilent Technologies 89410A Service Manual page 130

Vector signal analyzer
Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

Adjusting the Analyzer
To adjust input capacitance
2
Step 2. Set up the network analyzer.
a
Set the network analyzer as follows:
Receiver
Attenuation
Resolution BW
Source
Sweep Type
Sweep Mode
Sweep Time
Amplitude
Start Frequency
Stop Frequency
Display format
b
Connect the network analyzer's reference channel to A10 J2 using
an N(m)-to-BNC(f) adapter and SMB(f)-to-BNC(m) cable.
c
Connect a 10 k
d
Connect a 50
resistor.
e
Connect the network analyzer's source to the 50
termination using an N(m)-to-BNC(f) adapter and a BNC cable.
3
Normalize the measurement.
a
Wait for a complete sweep on the network analyzer.
b
Store the trace data in a storage register.
c
Set the network analyzer to divide the reference channel by the
storage register.
2-14
reference channel
20 dB
1 kHz
linear frequency sweep
continuous
1 s
-22 dBm
10 kHz
1 MHz
log magnitude
series resistor to the Agilent 89410A's input.
feedthrough termination to the 10 kW series
feedthrough