To Place The A82 In The Test Position - Agilent Technologies 89410A Service Manual

Vector signal analyzer
Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

Replacing Assemblies

To place the A82 in the test position

To place the A82 in the test position
1
Remove the front panel (see ''To remove front
panel''page
3-10).
3
Using a 2 point posidrive, remove the screw from both end
caps of the right side strap handle. Remove the strap
handle and side cover.
3-12
2
Using a T-15 torx driver, remove the screw from the back of
the top cover. Slide the cover off.
4
Using a T-15 torx driver, remove the four screws from the
power supply shield. Remove the power suppply shield.