Test Location
Signal Name
A91 P6 pin 10
CLK25M
A91 P6 pin 11
FSA2
A91 P6 pin 16
FSA1
A91 P1 pin 16
GTCLK
A91 P1 pin 22
PTA
A91 P1 pin 23
TPSD
A91 P1 pin 26
nSTRIGa
e
If the signals in the table are correct, the A36 Trigger is probably
faulty.
Troubleshooting the Analyzer
To troubleshoot trigger failures
Probable Faulty Assembly
A60 Frequency Reference
A60 Frequency Reference
A60 Frequency Reference
A36 Trigger
A55/A56 Sample RAM
A36 Trigger
A36 Trigger
1-87