Adjusting the Analyzer
To adjust anti-alias filter
A10 Analog Input Component Locator, Front View
4
Normalize the measurement.
a
Press the
[ setup 2
b
Set the network analyzer as follows:
Receiver
Attenuation
Resolution BW
Source
Sweep Mode
Sweep Type
Sweep Time
Amplitude
Start Frequency
Stop Frequency
Display format
c
Wait for a complete sweep on the network analyzer.
d
Store the trace data in a storage register.
e
Set the network analyzer to divide the reference channel by the
storage register.
2-20
] softkey.
reference channel
0 dB
1 kHz
continuous
linear frequency sweep
1 s
-13 dBm
100 kHz
20.1 MHz