2
Repeat the sample RAM self test without the Digital Source assembly.
a
Set the power switch to off ( O ).
b
Pull the A30 Digital Source assembly out of the card nest about 1
inch.
c
Set the power switch to on ( l ).
d
When the power-up tests are finished, press the following keys:
]
[System Utility
[
auto cal off]
[more cal setup
]
[
auto zero cal off]
]
[Return
[
]
more
[
]
diagnostics
[
service functions]
1125
[
enter
]
[
]
test log
[
]
Return
[
]
functional tests
[
sample RAM
]
e
If the sample RAM self test fails or aborts, the A55/A56 Sample
RAM assembly is probably faulty.
f
If the sample RAM self test passes, the A30 Digital Source
assembly is probably faulty.
Troubleshooting the Analyzer
To troubleshoot self-test lockup failures