Advertisement

5
Adjust A10 L39, L40, and L41.
a
Press the
[ setup 3
b
Set the network analyzer's reference level 0 dB.
c
Set the reference position at 80% of the display and the scale to
0.5 dB/div.
d
Put the marker at 10.2 MHz.
e
Adjust A10 L39, L40, and L41 for less than 1 dB ripple from
100 kHz to 10.2 MHz using the flat-edge adjustment tool. L39,
L40, L41 Correctly Adjusted
] softkey.
Adjusting the Analyzer
To adjust anti-alias filter
2-21