3
Check the
+
65 Vdc power supply.
a
Set the power switch to off ( O ).
b
Place the front panel in the A82 test position. (see
place the A82 in the test position", on
c
Set the power switch to on ( l ).
d
Check for
65 Vdc
+
e
If the voltage is correct, continue with step 4.
page
10% at TP3 using a Voltmeter.
±
Troubleshooting the Analyzer
To troubleshoot display failures
Chapter
3, "To
3-12.)
1-27