Vertical System Setup

To Change Vertical Reference Ground Position

You can adjust the waveform vertically by pressing the F3 softkey. The
following Figure 3- 3 displays the vertical position adjustment setting.
Figure 3-3 Adjust vertical position for the waveform displayed.
1 Press the F3 softkey to enable the setting of the waveform position.
2 As you turn the rotary switch
3 Notice that the waveform and the ground reference symbol moves
The maximum positive reference offset is 250 mV and the minimum negative reference
N O T E
offset is –250 mV.
28
Turn rotary switch clockwise to move the waveform up and vice- versa
to move the waveform down.
is displayed under the Position header shown in Figure 3- 3. This
voltage value indicates how far the ground reference is from the center
of the screen.
vertically with respect to the rotary switch position.
, the corresponding voltage value
Agilent U1602B/U1604B User's and Service Guide