Resistance Verification Test

1 Put the calibrator in Standby (STBY) mode.
2 Select Meter mode of the instrument, then press F2 softkey for
3 Press F1 to select resistance measurement.
4 Connect the calibrator to the instrument's input and ground terminal as
5 Select each function and range in the order shown in Table 7- 7. Apply
6 Compare the measurement results to the corresponding test limits
Table 7-7 Resistance Verification Test
Calibrator Settings
600 Ω
6 kΩ
60 kΩ
600 kΩ
3 MΩ
6 MΩ
60 MΩ
146
Ohmmeter.
shown in Figure 7- 2 using a banana plug.
the values from the calibrator shown in Table 7- 7 below.
shown in Table 7- 7.
Function
Range
600 Ω
2-Wire Ohms
6 kΩ
60 kΩ
600 kΩ
6 MΩ
6 MΩ
60 MΩ
Min Limit
595.8 Ω
5.958 kΩ
59.58 kΩ
595.8 kΩ
2.973 MΩ
5.958 MΩ
59.28 MΩ
Agilent U1602B/U1604B User's and Service Guide
Performance Tests
Max Limit
604.2 Ω
6.042 kΩ
60.42 kΩ
604.2 kΩ
3.027 MΩ
6.042 MΩ
60.72 MΩ
7