7
Performance Tests
Multimeter Performance Verification Tests
Use the Performance Verification Test to verify multimeter measurement
performance of the instrument. The performance verification tests are
based on the instrument's specification as listed in the "Chapter 8
Specification and Characteristic".
Some of the verification tests involve high voltages, hence only qualified personnel
WA R N I N G
is recommended to perform the tests.
To avoid electrical shock or personal injury, always place the calibrator in Standby
(STBY) mode before starting verification test.
DC Voltage Verification Test
1 Put the calibrator in Standby (STBY) mode.
2 Select Meter mode of the instrument, then press F1 softkey for Volt
3 Press F1 to select DC Voltage measurement.
4 Connect the calibrator to instrument's input and ground terminal as
5 Select each function and range in the order shown in Table 7- 5 and
6 Compare measurement results to the corresponding test limits shown in
Agilent U1602B/U1604B User's and Service Guide
Meter.
shown in Figure 7- 2 using a banana plug.
apply the values from the calibrator shown in Table 7- 5 below.
Table 7- 5.
143